Constellation, le dépôt institutionnel de l'Université du Québec à Chicoutimi

Effect of electrodeposition current density on the microstructure and the degradation of electroformed iron for degradable stents

Moravej Maryam, Amira Sofiene, Prima Frédéric, Rahem Ahmed, Fiset Michel et Mantovani Diego. (2011). Effect of electrodeposition current density on the microstructure and the degradation of electroformed iron for degradable stents. Materials Science and Engineering B: Solid-State Materials for Advanced Technology, 176, (20), p. 1812-1822.

[thumbnail of Effect of electrodeposition current density on the microstructure and the degradation of electroformed iron for degradable stents.pdf] PDF
Administrateurs seulement

3MB

URL officielle: http://dx.doi.org/doi:10.1016/j.mseb.2011.02.031

Résumé

Pure iron has become one of the most interesting candidate materials for degradable metallic stents due to its high mechanical properties and moderate degradation. In this work we studied the effect of electrodeposition current density on microstructure and degradation of pure iron films electrodeposited on Ti alloy substrate for degradable metallic stent application. Iron sheets were produced by electrodeposition using four different current densities 1, 2, 5 and 10 A dm-2. The films were then studied by SEM (scanning electron microscope) and EBSD (electron backscatter diffraction) to observe the surface morphology, grain size and orientation. Potentiodynamic polarization and static immersion tests were used to determine the corrosion rate and to study the degradation behavior of iron films, respectively. The current density was found to significantly influence the texture, the grain size and the grain shape of the electrodeposited iron. At current densities of 1, 5 and 10 A dm -2, weak textures corresponding to 〈1 0 1〉, 〈1 1 1〉 and 〈1 1 2〉 in the normal (electrodeposition) direction were obtained, respectively. At these current densities, average grain sizes smaller than 3 μm were also obtained. However, at 2 A dm-2, a strong 〈1 1 1〉//ND texture with density of 7.4 MUD was obtained with larger average grain size of 4.4 μm. The microstructure of iron samples changed after annealing at 550 °C because of the induced recrystallization. Different corrosion rates were obtained from potentiodynamic polarization curves of iron films deposited at different current densities because of their microstructures. Fe-2 showed the lowest corrosion rate due to its larger grains size and its texture. The corrosion rates of all iron samples decreased after annealing. Static degradation showed lower corrosion rates for iron because of the formation of a degradation layer on samples surface which reduced the degradation reaction rate. The degradation morphology was uniform for Fe-2 due to its strong texture. On the other hand, microscopic pits were found for Fe-1, Fe-5 and Fe-10.

Type de document:Article publié dans une revue avec comité d'évaluation
Volume:176
Numéro:20
Pages:p. 1812-1822
Version évaluée par les pairs:Oui
Date:2011
Sujets:Sciences naturelles et génie > Génie > Génie des matériaux et génie métallurgique
Département, module, service et unité de recherche:Départements et modules > Département des sciences appliquées > Module d'ingénierie
Mots-clés:Corrosion, Degradable stents, Electrodeposition, Electroforming, Microstructure, Texture
Déposé le:09 juin 2016 15:44
Dernière modification:09 déc. 2016 14:43
Afficher les statistiques de telechargements

Éditer le document (administrateurs uniquement)

Creative Commons LicenseSauf indication contraire, les documents archivés dans Constellation sont rendus disponibles selon les termes de la licence Creative Commons "Paternité, pas d'utilisation commerciale, pas de modification" 2.5 Canada.

Bibliothèque Paul-Émile-Boulet, UQAC
555, boulevard de l'Université
Chicoutimi (Québec)  CANADA G7H 2B1
418 545-5011, poste 5630